Noninferiority and Equivalence Evaluation of Clinical Performance among Computed Radiography, Film, and Digitized Film for Telemammography Services

Table 3

Mean values for the calculated variables by device.

VariableDeviceMeanSE⁢95% confidence interval
LowerUpper

TNCR0.470.050.360.57
ICR0.470.050.360.57
LUMIX0.460.050.360.57
FILM0.480.060.370.59

TPCR0.320.050.230.42
ICR0.300.050.210.39
LUMIX0.320.050.220.41
FILM0.350.050.250.45

FNCR0.150.030.080.21
ICR0.170.040.100.24
LUMIX0.160.030.090.22
FILM0.120.030.060.18

FPCR0.060.020.030.09
ICR0.060.020.030.09
LUMIX0.070.020.030.10
FILM0.050.020.020.08

AccuracyCR0.790.030.730.86
ICR0.770.040.700.84
LUMIX0.780.030.720.84
FILM0.820.030.760.89

PPVCR0.680.060.570.79
ICR0.680.060.560.79
LUMIX0.670.060.560.79
FILM0.680.060.570.79

SensitivityCR0.690.060.580.80
ICR0.640.060.520.76
LUMIX0.670.050.570.77
FILM0.740.060.630.84

SpecificityCR0.880.030.830.94
ICR0.880.030.830.94
LUMIX0.870.030.820.93
FILM0.900.030.850.96

AUCCR0.890.040.810.97
ICR0.870.050.770.96
LUMIX0.880.040.800.96
FILM0.900.020.870.94

Each mean was calculated from 490 observations (70 cases and seven radiologists). TN: true negative proportion, TP: true positive proportion, FN: false negative proportion, FP: false positive proportion, PPV: positive predictive value, ROC: receiver operating characteristic, AUC: area under ROC curve, SE: standard error of the mean.