Research Article
Remote ECG Monitoring by ECG247 Smart Heart Sensor
Table 2
ECG quality score
and the possibility to determine heart rhythm.
| | 12-lead ECG | ECG247 Smart Heart Sensor ECG | |
| Rhythm detection | 97 (100%) | 97 (100%) | 1.0 |
| P-wave () | 1.5 (0.7) | 1.1 (0.5) | <0.001 |
| QRS complex () | 1.0 (0.0) | 1.0 (0.0) | 1.0 |
| Noise artefacts () | 1.1 (0.2) | 1.3 (0.4) | <0.001 |
|
|
1: extremely good; 9: not accepted. |