Simultaneous Estimation of Metoprolol Succinate and Olmesartan Medoxomil in Pharmaceutical Formulation by Thin-Layer Chromatographic-Densitometric Method
Table 3
Summary of validation parameters of developed HPTLC method.
Validation parameters
MET
OLM
Linear range (g per band)
1000โ2250
800โ1800
LOD(a) (g per band)
249.61
210.05
LOQ(b) (g per band)
823.72
693.15
Accuracy (%)
99.85โ101.07%
99.08โ101.29%
Repeatability of measurement of peak area (RSD(c) %, )
0.1218
0.0742
Repeatability of sample application (RSD%, )
0.5823
0.2957
Precision (RSD %)
โIntraday
0.1547
0.0821
โInterday
0.1108
0.0442
(a)Limit of detection.
(b)Limit of quantification.
(c)Relative standard deviation. โ=โnumber of determinations.