Research Article
Implementation of Quality by Design for the Development and Validation of Pioglitazone Hydrochloride by RP-UPLC with Application to Formulated Forms
Table 6
Results of precision study.
| Intra-day precision () | Concentration injected (g mL−1) | Mean area ± SD | RSDa | Mean RT ± SD | RSDb |
| 100 | 222879 ± 1648 | 0.74 | 2.19 ± 0.004 | 0.16 | 200 | 502609 ± 590 | 0.12 | 2.19 ± 0.007 | 0.32 | 300 | 755267 ± 2276 | 0.30 | 2.20 ± 0.002 | 0.09 |
| Inter-day precision () | Concentration injected (g mL−1) | Mean area ± SD | RSDa | Mean RT ± SD | RSDb |
| 100 | 223992 ± 1130 | 0.50 | 2.19 ± 0.024 | 1.08 | 200 | 501345 ± 3070 | 0.61 | 2.19 ± 0.015 | 0.67 | 300 | 754190 ± 1555 | 0.21 | 2.18 ± 0.011 | 0.52 |
|
|
aRelative standard deviation based on peak area.
bRelative standard deviation based on retention time.
|