Research Article

Implementation of Quality by Design for the Development and Validation of Pioglitazone Hydrochloride by RP-UPLC with Application to Formulated Forms

Table 6

Results of precision study.

Intra-day precision ( )
Concentration injected
( g mL−1)
Mean area ± SDRSDaMean RT ± SDRSDb

100222879 ± 16480.742.19 ± 0.0040.16
200502609 ± 5900.122.19 ± 0.0070.32
300755267 ± 22760.302.20 ± 0.0020.09

Inter-day precision ( )
Concentration injected ( g mL−1)Mean area ± SDRSDaMean RT ± SDRSDb

100223992 ± 11300.502.19 ± 0.0241.08
200501345 ± 30700.612.19 ± 0.0150.67
300754190 ± 15550.212.18 ± 0.0110.52

aRelative standard deviation based on peak area.
bRelative standard deviation based on retention time.