Clinical Study

Reproducibility of Retinal Nerve Fiber Layer Measurements with Manual and Automated Centration in Healthy Subjects Using Spectralis Spectral-Domain Optical Coherence Tomography

Table 2

SD-OCT RNFL thickness reproducibility.

Manual centrationAutomatic centration
Mean RNFL in 𝜇 m (95% CI)SwCV %ICC (95% CI)Mean RNFL in 𝜇 m (95% CI)SwCV %ICC (95% CI)

Average98.35 (97.07–99.62)1.60.650.999 (0.998–0.999)98.57 (97.33–99.81)1.30.630.983 (0.974–0.990)
Temporal76.06 (74.59–77.53)2.80.970.980 (0.970–0.988)74.83 (73.54–76.11)1.10.860.998 (0.998–0.999)
Superior118.35 (116.36–120.31)5.30.830.997 (0.995–0.998)118.84 (116.95–120.73)2.10.800.999 (0.983–0.999)
Nasal71.80 (69.75–73.85)3.21.430.989 (0.984–0.994)72.86 (70.98–74.75)2.41.290.998 (0.998–0.999)
Inferior126.87 (124.83–128.91)2.50.800.997 (0.995–0.998)127.31 (125.35–129.27)1.90.770.999 (0.999–0.999)

SD-OCT: spectral-domain optical coherence tomography; RNFL: retinal nerve fiber layer; CI: confidence intervals; Sw: within subject standard deviation; CV: coefficient of variation; ICC: intraclass correlation.