Research Article

Development and Validation of Stability-Indicating RP-UPLC Method for the Determination of Methdilazine in Bulk Drug and in Pharmaceutical Dosage Form

Table 5

Results of precision study ( ).

Concentration injected, μg mL−1Intraday precisionInterday precision
Mean area ± SD% RSDa% RSDbMean area ± SD% RSDa% RSDb

20.0
1456560 ± 62770.430.311430424 ± 69480.480.32
40.0
2976574 ± 110580.370.242912796 ± 117690.400.28
60.04380383 ± 62280.140.134319865 ± 69870.160.17

aRelative standard deviation based on peak area.
bRelative standard deviation based on retention time.