Method Development and Validation for the Simultaneous Determination of Fexofenadine Hydrochloride and Montelukast Sodium in Drug Formulation Using Normal Phase High-Performance Thin-Layer Chromatography

Table 1

Linear regression data for calibration curves ( 𝑛 = 6 ).

ParametersFEXMTKT

Linearity range2400–10800 ng spot−1200–900 ng spot−1
Slope ± Standard error 1 . 6 5 1 ± 0 . 0 1 6 . 9 0 0 ± 0 . 0 4
Intercept ± Standard error 1 0 1 9 ± 9 0 . 2 7 1 6 . 1 2 ± 2 5 . 1 8
Confidence limit of slopea1.620 to 1.6826.796 to 7.003
Confidence limit of intercepta798.1 to 1240−77.73 to 45.49
𝑟 2 0.99960.9998
Sy.xb98.1927.38
P valuec<0.0001<0.0001

a95% confidence intervals.
bStandard deviation of residuals from line.
cP value is < 0.0001, considered extremely significant.