Research Article

High-Performance Liquid Chromatographic Method for Analysis of Emtricitabine in Rat Plasma: Method Development, Validation and Application to a Pharmacokinetic Study

Table 1

Intra-day and inter-day precision and accuracy of emtricitabine in rat plasma ( ).

Concentration (µg mL−1)Observed concentration
(µg mL−1)
Precision%
Accuracy%

Intraday
 0.10.097 ± 0.0033.09 97.0
 1.00.982 ± 0.0444.4898.20
 2.52.465 ± 0.0783.1698.0
Interday
 0.10.092 ± 0.0055.4592.0
 1.00.975 ± 0.063 6.4697.50
 2.5 2.427 ± 0.1054.32 97.08