Research Article

High-Performance Liquid Chromatographic Method for Analysis of Emtricitabine in Rat Plasma: Method Development, Validation and Application to a Pharmacokinetic Study

Table 2

Stability of emtricitabine in rat plasma ( ).

Sample conditionSpiked concentration (μg/mL)Mean determined concentration
(µg/mL)
Accuracy (%)

Bench-top stability0.10.09494.00
1.00.9898.00
2.52.4296.80

Freeze-thaw stability0.10.09999.00
1.01.03103.00
2.52.4698.40

One-week stability0.10.09292.00
1.00.8989.00
2.52.3895.20

Exposed at ambient temperature (25°C) for 4 h.
After three freeze-thaw cycles.
Stored at −16°C.