Scan Quality and Entrance Skin Dose in Thoracic CT: A Comparison between Bismuth Breast Shield and Posteriorly Centered Partial CT Scans

Table 2

Summary of findings from anthromorphic phantom study comparing partial CT with breast shield.

PositionNanoDot Avg Anterior (mrad)NanoDot Avg Posterior
(mrad)
Avg Effective Dose
(mAs)
CTDI
(mGy)
DLP
(mGy-cm)

Standard CT alone686602885.99318
Standard CT + breast shield426606875.93315
Partial CT590826956.47331

CTDI: CT dose index. DLP: Dose length product.