Research Article

Stability Indicating Simultaneous Validation of Telmisartan and Cilnidipine with Forced Degradation Behavior Study by RP-HPLC in Tablet Dosage Form

Table 5

Evaluation data of system suitability study for cilnidipine.

System suitability dataRSD (%)Theoretical platesAsymmetry
In-house limitNMTb 2.0NLTc 2000NMTb 2.0

Specificity0.26102141.35
Linearity0.6998231.23
Precision 0.14101951.29
Intermediate precision 0.22101661.40
Accuracy0.55100811.38
Solution stability0.29103721.15
Robustness0.62102241.26

Means not more than and means not less than.