Clinical Study

Changes in the Bispectral Index in Response to Experimental Noxious Stimuli in Adults under General Anesthesia

Table 4

Sensitivity and specificity at different BIS values.

CriterionSensitivity95% CISpecificity95% CI

≥16100.0088.4–100.00.000.0–11.6
>1696.6782.8–99.90.000.0–11.6
>1996.6782.8–99.93.330.08–17.2
>2486.6769.3–96.23.330.08–17.2
>2583.3365.3–94.410.002.1–26.5
>2680.0061.4–92.320.007.7–38.6
>2780.0061.4–92.323.339.9–42.3
>2873.3354.1–87.723.339.9–42.3
>2970.0050.6–85.326.6712.3–45.9
>3170.0050.6–85.333.3317.3–52.8
>3456.6737.4–74.533.3317.3–52.8
>3656.6737.4–74.536.6719.9–56.1
>3850.0031.3–68.740.0022.7–59.4
>4150.0031.3–68.756.6737.4–74.5
>4243.3325.5–62.663.3343.9–80.1
>4340.0022.7–59.466.6747.2–82.7
 >45*40.0022.7–59.473.3354.1–87.7
>4626.6712.3–45.980.0061.4–92.3
>4726.6712.3–45.986.6769.3–96.2
>4823.339.9–42.390.0073.5–97.9
>4916.675.6–34.790.0073.5–97.9
>5016.675.6–34.793.3377.9–99.2
>5113.333.8–30.793.3377.9–99.2
>5313.333.8–30.796.6782.8–99.9
>546.670.8–22.1100.0088.4–100.0
>580.000.0–11.6100.0088.4–100.0

Sensitivity and specificity of BIS in distinguishing moderate from severe noxious stimuli at different thresholds of BIS (criterion) with 95% confidence intervals.
>: superior; <: inferior; CI: confidence interval.
*Best cut-off ROC score. Threshold associated with maximization of the sums of sensitivity and specificity.