Evaluation of Stability of Rotating Hinge Knee Prostheses: A Biomechanical Study

Table 2

Results of the biomechanical analysis using the custom made biomechanical apparatus (Figures 3(a)3(c)).

ManufacturerStem length (mm)Stem taper (deg)Polyethylene inlay (mm)Minimum distraction to dislocate (mm)AL at 25 mm of distraction (deg)/method 1AL at 25 mm of distraction (deg)/method 2

Stryker—GMRS47010383,2 (SD: 0,33)3,3 (SD: 0,61)
DePuy—LPS/M.B.T.4602711,2 (SD: 0,80)11,3 (SD: 1,28)
DePuy—S-ROM Noiles4652618,4 (SD: 1,73)17,0 (SD: 1,13)
PLUS Orthopedics—RT-Plus3808303,0 (SD: 0,90)2,9 (SD: 0,90)
Zimmer—NexGen46012362,1 (SD: 0,68)2,3 (SD: 0,66)
Zimmer—NexGen60026422,3 (SD: 0,32)2,6 (SD: 0,44)

AL: angular laxity; SD: standard deviation.