Research Article

Quality by Design Approach for the Development and Validation of Glipizide, an Antidiabetic Drug, by RP-UPLC with Application to Formulated Forms and Urine

Table 5

Results of precision study.

Concentration injected (µg mL−1)Mean area ± SDRSDaMean Rt ± SDRSDb

Intraday precision ( )
1001165237 ± 23330.202.13 ± 0.0060.26
2002347961 ± 167280.712.13 ± 0.0060.28
3003482365 ± 231720.672.13 ± 0.0020.10

Concentration injected (µg mL−1)Mean area ± SDRSDaMean RT ± SDRSDb

Interday precision ( )
1001161904 ± 35350.302.14 ± 0.0180.82
2002347961 ± 167280.712.13 ± 0.0060.29
3003488820 ± 78500.232.14 ± 0.0200.91

aRelative standard deviation based on peak area.
bRelative standard deviation based on retention time.