Research Article

Rapid Tests versus ELISA for Screening of HIV Infection: Our Experience from a Voluntary Counselling and Testing Facility of a Tertiary Care Centre in North India

Table 2

Diagnostic performance of SD Bioline HIV-1/2 3.0 in comparison to Microlisa-HIV (reference standard).

Test 1 (SD Bioline HIV-1/2 3.0)Microlisa-HIV
ReactiveNonreactive

Reactive315
Nonreactive9742

HIV: human immunodeficiency virus.