Research Article

Flow Injection Potentiometric Assay of Hexoprenaline in Its Pure State, Pharmaceutical Preparations, and Biological Samples

Table 6

Statistical treatment of data obtained for the determination of Hexoprenaline using Hx conventional electrodes in comparison with the official method.

Hx-PTAHx-PMA

BatchFIABatchFIA

Pure solutions
X ± S . E . 1 0 0 . 2 3 ± 0 . 8 5 9 9 . 8 5 ± 0 . 7 4 9 9 . 8 0 ± 0 . 9 3 1 0 1 . 2 7 ± 0 . 2 1
Relative error (%)0.130.150.070.22
𝐹 3 , 3 value ( 9 . 2 7 ) 2.562.983.454.21
Slope of regression line0.9980.9860.9950.978
Intercept of regression line−0.087−0.0730.045−0.062

A s m a d o l 0 𝑥 0 0 0 𝑎 𝑒
X ± S . E . 1 0 0 . 5 4 ± 0 . 2 3 1 0 0 . 7 5 ± 0 . 7 4 9 9 . 6 3 ± 0 . 4 4 1 0 0 . 8 2 ± 0 . 5 0
Relative error (%)0.050.130.090.24
𝐹 3 , 3 value ( 9 . 2 7 ) 3.784.122.682.49
Slope of regression line0.9990.9880.9860.993
Intercept of reg. line0.012−0.023−0.0310.009

G y n i p r a l 0 𝑥 0 0 0 𝑎 𝑒
X ± S . E . 1 0 0 . 8 1 ± 0 . 3 4 1 0 0 . 2 5 ± 0 . 6 2 1 0 0 . 2 6 ± 0 . 4 8 1 0 1 . 2 ± 0 . 2 1
Relative error (%)0.140.520.180.31
𝐹 3 , 3 value ( 9 . 2 7 ) 4.573.512.883.69
Slope of regression line0.9920.9890.9960.985
Intercept of regression line−0.0910.0360.0180.042

Urine
X ± S . E . 1 0 0 . 5 4 ± 0 . 4 1 9 9 . 4 8 ± 0 . 5 6 1 0 0 . 5 6 ± 0 . 4 5 1 0 0 . 1 2 ± 0 . 3 4
Relative error (%)0.090.120.240.06
𝐹 3 , 3 value ( 9 . 2 7 ) 3.743.924.535.26
Slope of regression line0.9910.9840.9750.994
Intercept of regression line0.029−0.015−0.026−0.035

Plasma
X ± S . E . 1 0 0 . 2 ± 0 . 8 5 9 9 . 8 ± 0 . 7 4 9 9 . 8 ± 0 . 9 3 1 0 0 . 3 6 ± 0 . 5 4
Relative error (%)0.180.140.250.32
𝐹 3 , 3 value ( 9 . 2 7 ) 2.873.193.572.96
Slope of regression line0.9990.9780.9820.990
Intercept of regression line−0.014−0.0260.0340.091

X ± S . E . : average ± standard error.
O n e tailed critical 𝐹 -value. N.B.: for the official method (X ± S.E.) = 1 0 0 . 8 ± 0 . 6 5 .