Research Article

A Simple Automated Method for the Determination of Nitrate and Nitrite in Infant Formula and Milk Powder Using Sequential Injection Analysis

Table 3

Figures of merit. LD = detection limit, LQ = quantification limit, m = sensitivity (slope of calibration curve). LD and LQ calculated for real samples.

NitrateNitrite
ParameterNitrate (NO3-N)Nitrite (NO2-N)

Limits, solution
LD (mg L−1), (3  𝜎 /m, 𝑛 = 1 0 )0.170.025
LQ (mg L−1), (10  𝜎 /m, 𝑛 = 1 0 )0.550.084
Limits, real samples
LD (mg kg−1), (3  𝜎 /m, 𝑛 = 5 )0.320.05
LQ (mg kg−1), (10  𝜎 /m, 𝑛 = 5 )1.080.17
Precision (sr (%), 𝑛 = 5 )1.5–3.00.8–1.6
Linearity range (mg L−1)0.55–15.0 ( 𝑟 2 > 0 . 9 9 9 )0.084 - 1.01 ( 𝑟 2 > 0 . 9 9 9 )
Sampling rate (hour−1)3080