Research Article

Development of a Low-Cost SIA-Based Analyser for Water Samples

Table 1

Analytical figures of merit for the automated methods developed for chloride, nitrite, and nitrate in water samples. LD, LQ: detection (3σ) and quantification (10σ) limits. Precision was measured as repeatability ( 𝑛 = 1 0 ).

Analyte/Figure of meritLD
( 𝜇 g L−1)
LQ
( 𝜇 g L−1)
Linear range
( 𝜇 g L−1)
Precision 𝑠 𝑟
(%)
Difference with reference method (%)Sampling rate (hour−1)

Chloride0.61.901.9–541.42.060
Nitrite0.020.0760.11–3.82.9−2.860
Nitrate0.51.61.6–542.4−0.320