Research Article
Dual Wavelength RP-HPLC Method for Simultaneous Determination of Two Antispasmodic Drugs: An Application in Pharmaceutical and Human Serum
Table 4
Robustness of the method.
| Chromatographic conditions | Variation | Retention time (minutes) | PG | TMP |
| Temperature (°C) | 23 | 3.099 | 6.346 | 27 | 3.217 | 6.381 |
| Flow rate (mL/min.) | 0.9 | 3.341 | 6.413 | 1.1 | 2.89 | 5.994 |
| Vol. of acetonitrile (%) | 48 | 3.397 | 6.499 | 52 | 2.982 | 5.619 |
|
|