Robustness of the method.

ParametersVarianceAmount added (μg/mL)Retention time (min)Mean area of peak ± %RSD, = 3
StandardSample

Flow rate0.9 mL/min102.8219467 ± 0.3225203 ± 0.4
1.1 mL/min102.2184626 ± 0.3188516 ± 0.3
Organic (%) in mobile phase59102.7207066 ± 0.5206049 ± 0.3
61102.5203858 ± 0.3203051 ± 0.4
Detector wavelength278 nm102.7199421 ± 0.2198228 ± 0.6
282 nm102.7213519 ± 0.4211768 ± 0.7