Quantitative Determination of Lercanidipine Enantiomers in Commercial Formulations by Capillary Electrophoresis

Table 5

Linearity, LOD and LOQ, precision, and accuracy of the method for LER enantiomer analysis.

Limits(S)-LER(R)-LER

LOD (g/mL)0.80.6
LOQ (g/mL)2.41.9

Slope (RSD) 5.81.0

Intercept (RSD) 12.430.5

Linearity

Linear equation
Correlation coefficient ()0,9980,995

Precision and accuracy

Nominal concentration (g/mL)12.55010012.550100
Within-day ()
 Analyzed concentration (g/mL)12.449.8101.112.451.2100.1
 Precision (RSD, %)b3.44.43.14.53.73.2
 Accuracy (, %)c−0.8−0.41.1−0.82.40.1
Between-day ()
 Analyzed concentration (g/mL)12.351.798.612.352.198.4
 Precision (RSD, %)b3.64.34.04.34.03.8
 Accuracy (, %)c−1.63.4−1.4−1.64.2−1.6

Calibration graphs were prepared in triplicate () for concentrations of 12.5, 25, 50, 75, and 100 g/mL; , where is the peak height of LER enantiomers, is the slope, is the intercept, and is the concentration of the measured solution in g/mL; : number of determinations; bRSD (%): relative standard deviation expressed as percentage; (%): relative error expressed as percentage; : number of days.