Research Article

Determination of Levetiracetam in Human Plasma by Dispersive Liquid-Liquid Microextraction Followed by Gas Chromatography-Mass Spectrometry

Table 5

Within- and between-days accuracy and precision.

AnalyteNominal concentration (µg mL−1)Measured concentration (µg mL−1)Precision RSD (%)Accuracy RE (%)

Within-day
LEV2.01.910.6−3.1
4.04.111.43.1
40.038.07.6−5.0
60.057.88.9−3.6

Between-day
LEV2.02.04.61.6
4.04.2−8.15.1
40.039.13.0−2.4
60.061.64.02.6

RSD, relative error deviation expressed as percentage (%).