Research Article

Development of an Analytical Protocol for Determination of Cyanide in Human Biological Samples Based on Application of Ion Chromatography with Pulsed Amperometric Detection

Table 1

Parameters of the IC-PAD method for the determination of cyanide ion.

SampleLinear range [µg/L]Curve patternSDCV (%)Recovery [%]LOD [µg/L]

Urine1–100.9830.0010.42621.8
1–1000.9920.0031.6380
Saliva5–100.9880.0072.481045.1
5–1000.9940.051.84113
Sweat3–100.9870.0022.571085.8
3–1000.9930.010.8888