Research Article
Comparative Liquid Chromatographic Study for Concurrent Determination of Canagliflozin and Metformin in Combined Tablets
Table 2
Results obtained by the HPLC-UV and UPLC-UV methods for the simultaneous determination of CANA and MET in their binary mixture.
| Parameters | HPLC | UPLC | CANA | MET | CANA | MET |
| Linearity range (µg·mL−1) | 1–50 | 0.5–90 | 0.1–50 | 0.25–100 | Slope | 0.9102 | 0.9319 | 0.8821 | 1.0795 | Intercept | 0.3367 | 0.255 | 0.1499 | 0.5372 | Correlation coefficient | 0.9999 | 0.9999 | 0.9999 | 0.9999 | Accuracy (mean ± SD) | 99.81 ± 0.73 | 99.37 ± 0.54 | 99.47 ± 1.034 | 99.73 ± 0.89 | LOD | 0.7154 | 0.9327 | 0.463 | 0.614 | LOQ | 2.1075 | 2.8263 | 1.406 | 1.862 | Precision (% RSD) Repeatability | 1.31 | 1.47 | 0.48 | 0.71 | Interday precision (reproducibility) | 1.75 | 0.90 | 0.89 | 1.85 | Specificity (mean ± SD) | 99.60 ± 0.33 | 100.38 ± 1.20 | 99.89 ± 1.39 | 99.09 ± 1.31 |
|
|