Research Article
Development and Validation of an HPLC Method for Determination of Antidiabetic Drug Alogliptin Benzoate in Bulk and Tablets
Table 5
Robustness data of the proposed HPLC method.
| Parameter | %RSD of standard peak area | %RSD of assay |
| Column temperature | 25°C | 0.07 | 0.15 | 30°C (normal) | 0.03 | 0.19 | 35°C | 0.04 | 0.2 |
| Wavelength | 274 nm | 0.06 | 0.07 | 277 nm (normal) | 0.03 | 0.19 | 280 nm | 0.06 | 0.17 |
| Mobile phase composition | −5% acetonitrile | 0.05 | 0.20 | Normal | 0.03 | 0.19 | +5% acetonitrile | 0.02 | 0.14 |
| Flow rate | 0.8 ml/min | 0.04 | 0.11 | 1 ml/min (normal) | 0.03 | 0.19 | 1.2 ml/min | 0.08 | 0.23 |
|
|