Research Article

Analysis of Polycyclic Aromatic Hydrocarbons in Ambient Aerosols by Using One-Dimensional and Comprehensive Two-Dimensional Gas Chromatography Combined with Mass Spectrometric Method: A Comparative Study

Table 4

Limits of detection and quantification and recoveries of 18 PAHs obtained by the compared methods.

CompoundLODa (ng)LOQb (ng)Recovery ± RSD (%)
GC-qMSGC×GC-TOFMSGC-qMSGC×GC-TOFMSGC-qMSGC×GC-TOFMS

Nap0.070.400.211.1994.4 ± 4.2135 ± 45
Acy0.170.070.510.22119 ± 12116 ± 15
Ace0.050.170.160.52105 ± 5.3105 ± 7.8
F0.040.150.130.44158 ± 28130 ± 29
Phe0.100.340.311.0394.5 ± 5.386.3 ± 16
Ant0.190.310.580.9290.4 ± 4.695.1 ± 20
Fla0.050.140.160.4190.3 ± 3.9105 ± 13
Pyr0.080.360.251.0997.4 ± 5.397.2 ± 13
BaA0.120.090.370.2793.4 ± 4.986.9 ± 8.2
Chr0.040.080.130.2495.8 ± 5.8101 ± 16
BbF0.050.180.150.5396.1 ± 5.792.3 ± 10
BkF0.090.350.281.0594.2 ± 6.5105 ± 12
BeP0.130.130.400.3892.6 ± 5.892.7 ± 5.7
BaP0.120.240.370.7293.6 ± 5.3104 ± 9.0
Per0.110.340.321.0293.0 ± 5.592.5 ± 8.6
IP0.150.650.161.9495.0 ± 5.493.9 ± 8.5
DBahAnt0.131.050.403.1494.9 ± 5.595.8 ± 5.7
BghiPer0.090.220.270.6694.6 ± 6.087.0 ± 8.5

aLOD, smallest amount of analyte that is statistically different from the blank; bLOQ, smallest amount of analyte that can be measured with reasonable accuracy.