Optimization and Validation of Thermal Desorption Gas Chromatography-Mass Spectrometry for the Determination of Polycyclic Aromatic Hydrocarbons in Ambient Air

Table 5

Soxhlet extraction-GC/MS method validation parameters for the 16 EPA PAHs in SRM 1649b ().

PAHExperimental mean (ng)aNIST-certified value (ng)aRSD (%)Recovery (%)Accuracyb (%)

Naph67.3 ± 18.6391 ± 3533.817.2−82.8
Acy20.2 ± 5.979.9 ± 9.535.925.2−74.8
Ace25.0 ± 5.481.6 ± 16.526.530.7−69.4
FL32.6 ± 9.292.3 ± 14.434.535.3−64.7
Phe1215 ± 3311668 ± 2433.472.8−27.2
Ant104 ± 25169 ± 132.561.8−38.2
Ft2392 ± 5592573 ± 3231.093.0−7.05
Pyr1914 ± 3982054 ± 5727.693.2−6.79
BaA808 ± 148870 ± 2024.392.9−7.13
Chry1632 ± 4641256 ± 1137.612929.9
BbFt + BjFt3144 ± 8003260 ± 9133.796.4−3.58
BkFt921 ± 319702 ± 2045.913131.3
BaP1019 ± 2711018 ± 9835.21000.07
IP1109 ± 2961192 ± 6535.493.1−6.89
DBahA + DBacA507 ± 176241 ± 445.9363263
BghiP2143 ± 5801777 ± 3235.812020.6
Averagec34.9120|36.8|

aExpanded uncertainty about the mean, with coverage factor, k = 2; baccuracy = (experimental value − certified value) × 100/certified value; cexcept Naph, Acy, Ace, and FL.