Journal of Analytical Methods in Chemistry / 2018 / Article / Tab 1 / Research Article
Speciation Analysis of Arsenic Compounds by HPLC-ICP-MS: Application for Human Serum and Urine Table 1 Operating conditions of anion-exchange chromatography (AEC) combined with inductively coupled plasma mass spectrometry (ICP-MS) for speciation analysis of arsenic compounds.
Anion-exchange chromatography operating conditions Pump Quaternary pump LC 10 A (Shimadzu, Japan) including online degasser, column oven, and system control unit Column Hamilton PRP-X100 anion-exchange column (250 × 2.1 mm × 10 µ m particle size) with an associated guard column Column temperature 30°C Mobile phase Chanel A: 5 mM (NH4 )2 CO3 , pH 9.0, 0.05% Na2 EDTA Chanel B: 50 mM (NH4 )2 CO3 , pH 9.0, 0.05% Na2 EDTA Chanel C: MeOH Elution mode gradient Flow rate of mobile phase 400 µ L·min−1 Injection volume 100 µ L ICP-MS operating conditions ICP-MS system ELAN 9000 (PerkinElmer Sciex, Penlivia, Canada) RF powder 1300 W Plasma gas flow 16 L·min−1 Auxiliary gas flow 1.25 L·min−1 Nebulizer gas flow 0.9 L·min−1 Nebulizer Cross-flow Spray chamber PFA double pass Monitored ion As+ (m /z 74.92) Dwell time 100 ms Measurement mode Peak hopping