Regression and statistical parameters obtained from differential pulse voltammetry calibration curves of naproxen and its degradation product using SWCNTs/CPE.

ParametersNAPDegradation product
At 0.85 VAt 1.18 V

Concentration range (µg·mL−1)0.067–0.670.067–1.000.067–0.73
Slope of regression line (μA)0.90831.22640.7696
sa (μA)0.01540.03210.0096
Intercept of regression line (b) (μA·mL·μg−1)0.0021−0.07790.0141
sb (μA·mL·μg−1)0.006260.01940.0043
Correlation coefficient (r)0.99880.99560.9993
LOD (µg·mL−1)0.10420.08120.045
LOQ (µg·mL−1)0.315720.24610.135
SD0.0230.0250.015
RSD (%)3.042.783.46