Investigating Advanced Approaches Based on Iso-Absorptivity Coefficient in Unresolved Spectral Signals of Binary Mixtures
Table 2
Assay parameters and validation sheet obtained by applying the proposed spectrophotometric methods.
Parameters
a-Centering
AS
AM
A-Sum
OFX
DXM
OFX
DXM
OFX
DXM
OFX
DXM
Wavelength (nm)
D0 at 298
D0 at 239.5
D0 at 238
P at 238 nm
D1 at 246.5 nm
Calibration rangea (µg/mL)
1–10
3–10
3–10
3–10
2–10
Slope
0.1154
0.0382
0.0387
1.0012
0.0087
Intercept
−0.0084
−0.0008
−0.0008
−0.0011
−0.0017
Correlation coefficient (r)
0.9999
0.9998
0.9998
0.9999
0.9999
Meana
99.90
99.52
100.17
99.89
100.53
RSD%
0.666
0.515
0.931
0.449
0.516
Accuracyab
100.78 ± 1.061
99.97 ± 0.687
100.32 ± 1.002
99.55 ± 0874
99.32 ± 0.417
100.11 ± 0.744
100.54 ± 0.987
99.88 ± 0.552
Repeatabilityac
±0.56/0.559
±0.68/0.684
±0.47/0.474
±0.96/0.965
±0.63/0.633
±0.41/0.411
±0.54/0.541
±0.97/0.965
Interday precisionac
±0.70/0.698
±0.71/0.711
±0.55/0.552
±1.11/1.105
±0.73/0.729
±0.63/0.632
±0.71/0.711
±0.97/0.977
Robustnessacd
100.06/0.785
99.23/0.884
100.33/0.877
99.52/0.987
99.99/0.719
100.98/1.050
100.41/0.963
99.87/1.114
System suitability RSD%e
0.774
0.574
0.411
0.217
0.419
0.622
0.796
0.433
aAverage of three experiments. bMean ± standard deviation of 3 concentrations of each drug (4, 6, and 8 µg/mL). cStandard deviation/relative standard deviation of 3 concentrations of each drug (4, 6, and 8 µg/mL). dRobustness was checked by testing the effect of the solvent (29, 31, and 32% methanol). eOFX and DXM (6 µg/mL).