Research Article

Fingerprint Spectrophotometric Methods for the Determination of Co-Formulated Otic Solution of Ciprofloxacin and Fluocinolone Acetonide in Their Challengeable Ratio

Table 2

Assay parameters and validation sheets for pure cited drugs at their maxima.

ParameterCIPFLU

Wavelength (nm)278.0238.0
N77
Range (μg/mL)3.0–15.03.0–15.0
Intercept−0.0007−0.0011
Slope0.12090.0357
Correlation coefficient0.99990.9999
Accuracya,b100.08 ± 0.5199.79 ± 0.54
Repeatabilitya,c0.430.82
Interday precisiona,c1.191.03

a-Centering, EXRS, CM, UCS, and SS. a-Centering and RS. aAverage of three experiments. bMean ± standard deviation. cRSD% of concentrations (4.0, 8.0, and 12.0 μg/mL of both CIP and FLU).