Research Article

Lipemia Interferences in Biochemical Tests, Investigating the Efficacy of Different Removal Methods in comparison with Ultracentrifugation as the Gold Standard

Table 3

Comparison of the results of native serum, serum blank, and diluted serum with ultracentrifuged serum.

ParameterModerate lipemia (TG: 700–1000 mg/dl)Allowable bias (%)
Native serum (%)Serum blank (%)Serum dilution (%)

ALT+1011.48
Albumin+70+341.43
ALP06.72
Amylase+77.4
AST06.54
Bilirubin (total)08.95
Calcium+32+20+340.82
Chloride+35+6+330.5
CPK−13−11−2411.5
Creatinine−70−343.96
GGT−30−2+14411.6
Glucose+160+332.34
Iron+11+26+418.8
LDH+24.3
Lipase+42−20+17711.31
Magnesium+146+92+1231.8
Phosphorus+106+103+1283.38
Total protein+26−8+321.36
TIBC+13+16+341.3
Urea−6−6−215.57
Uric acid+24.87

Difference with ultracentrifuged serum, calculated as bias.