Lipemia Interferences in Biochemical Tests, Investigating the Efficacy of Different Removal Methods in comparison with Ultracentrifugation as the Gold Standard

Table 4

Comparison of the results of native serum, serum blank, and diluted serum with ultracentrifuged serum.

ParameterSevere lipemia (TG > 1000 mg/dl)Allowable bias
Native serum (%)Serum blank (%)Serum dilution (%)

ALT011.48
Albumin+130+431.43
ALP−36.72
Amylase+14+4+157.4
AST+14−27+366.54
Bilirubin (total)08.95
Calcium+40+30+570.82
Chloride+82+22+900.5
CPK−22−5−6111.5
Creatinine−20−8003.96
GGT−911.6
Glucose+44+2+632.34
Iron−9+660+218.8
LDH+9+8+74.3
Lipase+711.31
Magnesium+180+108+3251.8
Phosphorus+109+187+2103.38
Total protein+59−3+701.36
TIBC+34+33+781.3
Urea+11+17−175.57
Uric acid+44.87

Difference with ultracentrifuged serum, calculated as bias.