Research Article

Selective Consecutive Determination of Desloratadine and Montelukast Sodium in Their Pure and Binary Dosage Form Based on Pencil Graphite Electrochemical Sensors

Table 4

Determination of DES and MON in laboratory prepared mixtures by the proposed sensors.

RatioRecovery %
DESMONDES.RNK sensorMON.Cd. sensorCombined sensor
DESMON

11100.65100.98100.54100.92
1299.93100.5699.8999.75
1598.23100.2398.11100.05
21100.1899.34100.0899.22
51100.8798.92100.76100.79
Mean ± SD99.97 ± 1.04100.01 ± 0.8699.71 ± 1.30100.15 ± 0.71