Determination of Finasteride in Tablets by High Performance Liquid Chromatography

K. Basavaiah and B. C. Somashekar

Department of Post Graduate Studies and Research in Chemistry, University of Mysore, Manasagangotri, Mysore-570006, India

Received 1 September 2006; Revised 22 October 2006; Accepted 2 November 2006

Copyright © 2007 Hindawi Publishing Corporation. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

A rapid, highly sensitive high performance liquid chromatographic method has been developed for the determination of finasteride(FNS) in bulk drug and in tablets. FNS was eluted from a ODS C18 reversed phase column at laboratory temperature (30 ± 2°C) with a mobile phase consisting of methanol and water (80+20) at a flow rate of 1 mL min-1 with UV detection at 225 nm. The retention time was ∼ 6.1 min and each analysis took not more than 10 min. Quantitation was achieved by measurement of peak area without using any internal standard. Calibration graph was linear from 2.0 to 30 μg mL-1 with limits of detection (LOD) and quantification (LOQ) being 0.2 and 0.6 μg mL-1, respectively. The method was validated according to the current ICH guidelines. Within-day co efficients of variation (CV) ranged from 0.31 to 0.69% and between-day CV were in the range 1.2-3.2%. Recovery of FNS from the pharmaceutical dosage forms ranged from 97.89 – 102.9 with CV of 1.41-4.13%. The developed method was compared with the official method for FNS determination in its tablet forms.