Research Article

A Simple and Sensitive HPTLC Method for Simultaneous Determination of Metformin Hydrochloride and Sitagliptin Phosphate in Tablet Dosage Form

Table 3

Data indicating various validation parameters of HPTLC method ( ).

METSITA

Precision ( )
Mean ± SD; RSD (%)
 Method precision99.86 ± 0.88; 0.8898.91 ± 1.21; 1.23
 Intraday precision100.43 ± 0.49; 0.4999.12 ± 0.71; 0.72
 Interday precision100.46 ± 1.14; 1.13100.56 ± 1.43; 1.43
LOD (ng/μL)13.052.65
LOQ (ng/μL)39.568.03