Research Article

RP-HPLC Method for Simultaneous Estimation of Cefepime Hydrochloride and Tazobactam Sodium in Bulk and Pharmaceuticals

Table 2

Results of validation parameters.

ParameterCEFETAZO

System suitability parameters ( )
 Retention time (min) ± % RSD10.49 ± 0.687.69 ± 0.83
 Capacity factor ± % RSD3.19 ± 0.352.07 ± 0.84
 Theoretical plate ± % RSD10505 ± 1.1112957 ± 0.65
 Asymmetry factor ± % RSD1.02 ± 0.920.99 ± 1.24
Linearity study
 Linearity range (μg/mL)4–240.5–3.0
 Slope7373048802
 Intercept274772327.9
 Correlation coefficient ( )0.99770.9974
Precision ( )
 Repeatability (% RSD)0.921.09
 Interday (% RSD)0.270.54
 Intraday (% RSD)0.510.29
Accuracy ( )
 80% ± (% RSD)102.22 ± 0.2599.92 ± 1.09
 100% ± (% RSD)103.01 ± 0.93102.02 ± 0.76
 120% ± (% RSD)100.95 ± 0.6899.82 ± 1.34
Specificity studyPercent drug remained
 Acid stress95.9482.29
 Base stress94.9598.63
 Oxidation stress67.1987.4
 Thermal stress92.2891.27