Abstract

A green, simple, and stability-indicating RP-HPLC method was developed for simultaneous determination of permethrin isomers in pharmaceutical preparations. The separation was based on a C18 analytical column (150 × 4.6 mm, i.d., 5 μm). The mobile phase consisted of ethanol: phosphoric acid solution (pH = 3) (67 : 33, v/v). The elution was carried out at 30°C temperature with a flow rate of 1.0 mL/min. Quantitation was achieved with UV detection at 215 nm. In forced degradation studies, the drug was subjected to oxidation, hydrolysis, photolysis, and heat. The method was validated for specificity, linearity, precision, accuracy, and robustness. The applied procedure was found to be linear in permethrin concentration range of 0.5–50 μg/mL with correlation coefficients of 0.9996 for each isomer. Precision was evaluated by replicate analysis in which % relative standard deviation (RSD) values for areas were found below 2.0. The recoveries obtained (99.24%–100.72%) ensured the accuracy of the developed method. The peaks of permethrin isomers well resolved from various degradation products as well as the pharmaceutical excipients. Accordingly, the proposed validated and sustainable procedure was proved to be proper for routine analyzing and stability studies of permethrin in pharmaceutical preparations.

1. Introduction

Permethrin, (±)-3-phenoxybenzyl 3-(2,2-dichlorovinyl)-2,2-dimethylcyclopropanecarboxylate (Figure 1), is a synthetic mixture of cis- and trans-isomers of pyrethrin, chemically altered to bestow light and heat stability [1]. Permethrin isomers have different chemical, physical, and toxicological properties. It has been reported that the toxicity of permethrin with a cis- : trans- ratio of 25 : 75 is lower than that of permethrin with a cis- : trans- ratio of 40 : 60. This drug acts on the nervous system of insects and interferes with sodium channels to disrupt the function of neurons, and causes muscles to spasm, culminating in paralysis, and death [24]. Permethrin is used predominantly for treatment of scabies, caused by head lice and mites [5, 6]. Recently, studies have demonstrated the efficacy of permethrin therapy in the treatment of acne rosacea for its mite eradicating ability [710]. Moreover, it has been used as ectoparasiticides in agriculture, in the impregnation of clothing, and also in veterinary medicine, since pyrethroids development in the 1970s [11].

Concerning new therapeutic application of permethrin, validated and stability-indicating methods should be available to determine this drug in topical preparations. However, no method was compiled in British and United state pharmacopoeias for the analysis of this medicine in pharmaceutical preparations. Few methods reported the quantitation of permethrin in pharmaceutical dosage forms using spectrophotometric methods, [12] and LC methods with ultraviolet detection [1316], most of which have not been validated according to International Conference on Harmonization (ICH) stability-indicating guidelines.

In all of the aforementioned LC techniques, acetonitrile or methanol has been used as a part of mobile phase or extraction procedure. It is worth mentioning that these solvents are ranked by US Environmental Protection Agency (EPA) as hazardous solvents [17] and because of their inherent toxicity [18], safe detoxification of the waste solvents is essential, which may lead to high to very high disposal costs.

Due to scientific and public concern about the environment pollution, environmentally friendly practices have been introduced in different areas of society and research. In green analytical chemistry, sample preparation and LC analysis need special attention because hazardous solvents, are often used. Possibilities toward green LC include reducing solvent use, switching to more benign solvents and/or eliminating organic solvents [1923].

Taking ICH guidelines in consideration, the present study describes a simple, validated, and stability-indicating analytical method for determination of permethrin isomers, which meets the green aspects in analytical chemistry. The performances of the method were evaluated, and its potential for the determination of permethrin isomers in pharmaceutical preparations was investigated.

2. Experimental Section

2.1. Materials, Reagents, and Chemicals

Qualified permethrin standard (99.88%) was kindly provided by Gilaranco (Rasht, Iran). All solvents and reagents were of gradient and analytical grade, respectively, and were purchased from Merck (Darmstadt, Germany). HPLC grade water was obtained through a Milli-Q system (Millipore, Milford, MA, USA) and was used to prepare all solutions. The pharmaceutical formulations (cream 5%) and the corresponding placebos (mixture of all the excipients) were prepared in our laboratory. Permethrin cream 5% manufactured by Gilaranco (Rasht, Iran) was purchased from local pharmacy.

2.2. Equipments

The HPLC method was carried out on a Shimadzu HPLC system (Shimadzu, Kyoto, Japan), set to recycle the mobile phase and was equipped with an SCL-10AVP system controller, LC-10 ADVP pump, DGU-14A degasser, and a SPD-M10AVP PDA detector. The peak areas were integrated automatically by computer using a Shimadzu Class VP software program. A 20 μL volume of sample was introduced into a Rheodyne model 7725i injector.

2.3. Chromatographic Conditions

The elution was carried out on a C18 column (150 mm × 4.6 mm, 5 μm particle size) from Teknokroma (Barcelona, Spain). All analyses were performed at the column temperature of °C under isocratic conditions with a mobile phase of ethanol: phosphoric acid solution (pH = 3.0) (67 : 33, v/v), and a flow-rate of 1.0 mL/min, using PDA detection at 215 nm.

2.4. Preparation of Standard Solutions

Stock standard solution of permethrin was prepared in ethanol 96% at a concentration of 10 mg/mL. This solution was found to be stable for at least 1 month, when stored at 2–8°C. Freshly prepared working standards at concentration levels of 0.5, 2, 5, 10, 20, and 50 μg/mL were obtained from stock solution by the appropriate dilution in HPLC-grade water.

2.5. Preparation of Test Solutions

A 1 gr portion of cream (equivalent to 50 mg of permethrin) was transferred into a 100 mL volumetric flask. The volume was adjusted to the mark with ethanol 96% to provide a theoretical concentration of 500 μg/mL of permethrin. The solution was serial diluted with HPLC-grade water to make final concentration of 9 μg/mL. The experiment was performed in triplicate. These samples were assayed using calibration curves of working standard solutions. The same procedure was applied to placebo to be sure about the selectivity of the method.

2.6. Method Development

A variety of mobile phases were investigated in the development of a stability-indicating LC method for the analysis of permethrin isomers in pharmaceutical preparations. The suitability of mobile phase was decided on the basis of green analytical chemistry principles, selectivity, and sensitivity of the assay, stability studies, and separation of permethrin isomers from the impurities formed during forced degradation studies. Different wavelengths were also investigated.

2.7. Forced Degradation Studies

The stability-indicating capability of the method was determined by subjecting reference solutions (50 μg/mL) to accelerated degradation by acidic, basic, heat, oxidative, and photolytic conditions to evaluate the interferences in the quantitation of permethrin isomers. Sample solutions prepared in 1 M hydrochloric acid and 1 M sodium hydroxide were used for the acidic and basic hydrolysis, respectively. Both solutions were kept at ambient temperature for 12 h and then neutralized with basic or acidic solutions, as necessary. For evaluating the heat condition, the reference solution heated at 80°C for 12 h. For oxidative degradation, sample solutions were prepared in a solution of hydrogen peroxide (3%) and kept at ambient temperature for 4 h, protected from light. Photodegradation was induced by exposing the sample solution to UV-Lamp at a wavelength of 254 nm placed in a wooden cabinet for 1 hour. The solutions were diluted with HPLC-grade water to final concentration of 10 μg/mL and were injected into chromatograph.

2.8. Method Validation

The developed method was validated as per the requirements of the ICH guidelines. Linearity was evaluated by determining six working standard solutions at a concentration range of 0.5–50 μg/mL. Five sets of such solutions were prepared. Each set was analyzed to plot a calibration curve. Slope, intercept, and coefficient of determination () of the calibration curves were calculated to ascertain linearity of the method. The limit of quantification (LOQ) was defined as the lowest concentrations with the RSDs lower than 5% and accuracies within ±5%, considering at least seven times the response compared to that of the blank. In order to check the robustness, the effect of small but deliberate variations in the chromatographic conditions was evaluated. The conditions studied were flow rate (altered by ±0.2 mL/min), column temperature (altered by ±2°C), and pH of phosphoric acid solution (altered by ±0.2). These chromatographic variations were evaluated for resolution between permethrin isomers and % assay of the drug. For method repeatability, assay of working standard solutions (0.5, 10, 20 and 50 μg/mL) was repeatedly performed five times on the same day (intraday). For reproducibility, freshly prepared solutions at aforementioned concentration levels were analyzed at different days (interday), and results were statistically evaluated in terms of % RSD. For recovery studies, 0.5 gr portions of preassayed permethrin cream 5% were spiked with extra 1, 2, and 4 mL of stock standard solution. These samples were handled as explain in Section 2.5, and the final target levels of 7, 9, and 13 μg/mL were prepared. The concentrations were calculated using calibration curves. Accuracy was calculated as the deviation of the mean from nominal concentration. To assess accuracy, freshly prepared placebo of the permethrin pharmaceutical formulations was spiked with various amounts of permethrin to obtain the concentration levels of 0.5, 10, 20, and 50 μg/mL. Each solution was injected in triplicate. Selectivity of this method was indicated by the absence of any excipient interference at retention times of the peaks of permethrin isomers. The absence of interfering peak was evaluated by injecting a blank sample consisting of diluent and placebo. The double check of the lack of interferences of the resulting by-products with the elution of the peaks of permethrin isomers was done by calculating the F factor, meaning the ratio of the UV molar absorption coefficients of permethrin isomers at the 215 nm and 272 nm, respectively, using (1) where A (215) and A (272) are the permethrin peaks areas obtained at 215 nm and 272 nm, respectively. The resulting F factors were compared with those of the standards [24].

2.9. Estimation of the Uncertainty of the Measurements

An expanded uncertainty budget was constructed for permethrin isomers in pharmaceutical preparations by the RP-HPLC-DAD method according to previously reported procedures [2528].

Four individual sources were taken into account to assess the expanded uncertainty.

2.9.1. Uncertainty of the Measurement Standard

The uncertainty of the measurement standard is calculated by the quadratic addition of two terms: the uncertainty certified by manufacturer () and the uncertainty corresponding to its preparation by dilution or weighting () (2) [25]. The stock uncertainty () is calculated from a value given by the manufacturer using where the purity is expressed as P%. When there are independent standard preparations at each concentration level, the term could be eliminated. In this case, the contribution of this term is included in the term [25, 26].

2.9.2. Uncertainty Associated to the Calibration Curve

This value is calculated using (4) [27]. where is the concentration calculated from the calibration curve and is the standard deviation of the concentration, calculated from the calibration curve using where is the slope of the line, is the y-intercept of the line, is the number of replicate unknowns, is the number of the standards, is the mean of repeat measurements of for the sample, is the mean of the values for the calibration standards, are the concentrations of the standards, is the average concentration of the standards.

2.9.3. Uncertainty Associated to Precision

This value is calculated using where is standard deviation of the experimental data for precision and is the number of assays [27].

2.9.4. Uncertainty Associated to Accuracy

This parameter is calculated using is the relative standard deviation of the recovery and is the number of assays [27].

The value of the expanded uncertainty was calculated according to ISO GUM guidelines using where is expanded uncertainty, the coverage factor (for confidence interval 95%, ), is the concentration of the drug [28].

3. Results and Discussion

3.1. Optimization of the Chromatographic Conditions

Optimum conditions, which are necessary for the quantitative analysis of the permethrin isomers with maximum sensitivity, were established by a number of preliminary experiments. In this context, different detection wavelengths and compositions of mobile phase, considering the green chemistry principles [18] were explored. Optimum conditions were fixed by varying one parameter at a time while fixing other parameters constant and observing its effect on the peak resolution and also on the response.

Initially, ethanol was replaced with acetonitrile in mobile phase previously reported containing acetonitrile: water (70 : 30, v/v) [29], in order to use a benign and green solvent. Although ethanol has some shared or similar characteristics to methanol and acetonitrile, some properties are different. By nature, ethanol has twice the viscosity of methanol, and almost 4 times the viscosity of acetonitrile. In addition, it has different strength of the solvents. Thus, the mobile phase composition as well as the operating conditions needed to be optimized. The aforementioned change in mobile phase resulted in unresolved permethrin isomers appearing before 12 minutes. In an attempt to improve separation, ethanol content of mobile phase were decreased, and the resolution values of higher than 2 was achieved with the ratio of 65 : 35. But in this situation, the peak shapes and retention times were not satisfactory. Decreasing the pH of water to 3.0 using phosphoric acid solution helped to reduce the retention times and to sharpen the peaks. The chromatograms related to the above procedures are shown in Figure 2. According these preliminary results, the detection wavelength of 215 nm and the mobile phase of ethanol: phosphoric acid solution (pH = 3) (67 : 33, v/v) were finalized. Under such chromatographic condition, trans- and cis-isomers separated adequately and appeared at 16.2 and 20.3 min, respectively, (Figure 3(a)). Before fully implemented in the quantitative determination of drug substance and pharmaceutical preparation, this method was thoroughly validated according to ICH guidelines.

3.2. Forced Degradation Studies

Stability-indicating method is defined as an analytical method that accurately quantifies the active ingredients without interference from degradation products, process impurities, excipients, or other potential impurities [30]. The forced degradation studies in photolytic condition, resulted in nonsignificant decrease of the area without any detectable eluting degradation product (Figure 3(f)). However, under the acidic condition, permethrin isomers exhibited a significant decrease, in that only 68.08% of trans- and 64.57% of cis-isomers were recovered and four additional hydrophilic peaks were detected before 6 min (Figure 3(c)). The heat condition also exhibited decrease of the areas to 96.80% and 97.54% for trans- and cis-isomers, respectively, and presumably only one peak was detected, which eluted in solvent front (Figure 3(b)). Under the oxidative condition, the trans- and cis-isomer contents decreased to 88.05% and 84.76%, respectively, without any additional peak (Figure 3(e)), indicating that the degradation products did not elute in the HPLC method or may have been degraded to nonchromophoric products. As reported previously [15], permethrin isomers were degraded thoroughly under alkaline condition, and three additional peaks were observed before 6 min (Figure 3(d)). Summary of all degradation studies is mentioned in Table 1.

3.3. Method Validation
3.3.1. Selectivity and Specificity

The application of the whole procedure to placebo samples in order to verify the method selectivity demonstrates that no interferences were detected (Figure 4(b)). The specificity of the method was illustrated by the complete separation of permethrin isomers from each other and from different degradation products as shown in Figure 3. Furthermore, the decreases observed in isomer contents in stability studies, when degradation products appeared, proved the specificity of the method (Table 1). Also, the F factors of the permethrin isomers in stress solutions [(15.70–15.90), (15.43–15.74)] show the same values of that of the standards (15.80, 15.58) for trans- and cis-isomers, respectively.

3.3.2. Linearity, Precision and LOQ

Linearity was determined by constructing five independent calibration curves each one with six calibration points of permethrin, including the LOQ, in the range of 0.5–50 (sum of isomers) μg/mL. The peak area of each isomer against the respective concentration was used for plotting the graph considering the ratio of trans- : cis- as 3 : 1, and the linearity evaluated by the least square regression analysis. The linearity curves were defined by the following equation for trans-isomer: y = 225965x + 43623, and y = 220640x + 9149.9 for cis-isomer. The value of the determination coefficient calculated ( = 0.9996 for each isomer) indicated the linearity of the calibration curve for the method. Moreover, the relative standard error of slope can be used as a parameter with respect to the precision of the regression, as a general acceptance criterion for the linearity performance of the analytical procedure [31]. This parameter should be comparable to the calculated RSD in the evaluation of the precision. In this study, the results obtained for the RSDs of the slopes were 2.14% and 1.39% for trans- and cis-isomers, respectively, which are comparable to those of the maximum respective precision values of 2.07% and 1.94% (Table 2).

The results from the validation of method are summarized in Table 2. The method proved to be precise, as the intra- and interday precision ranged from 0.70%–1.90% and 1.30%–2.07%, respectively, for all analytes. These values fulfill the validation criteria of an analytical method designed for quality control of pharmaceutical preparations for which RSD values <2% are acceptable [31].

The LOQ is the lowest concentration that can be quantified with acceptable precision and accuracy. The LOQ of permethrin was determined to be 0.5 μg/mL, considering the mean accuracy value of 101.18% and maximum RSD value of 2.07% (Table 1). This value indicates that the proposed method is much more sensitive than what have been reported previously [1216].

3.3.3. Recovery and Accuracy

The accuracy was evaluated applying the proposed method to the analysis of the in-house mixture of the cream excipients with known amounts of the drug, to obtain solutions at concentrations of 0.5, 10, 20, and 50 μg/mL. The accuracy was assessed from three replicate determinations and calculated as the percentage of the drug recovered from the formulation matrix. The means and RSD% obtained for permethrin isomers are shown in Table 2 with a range of 99.53%–101.28% and 99.58%–101.42% for trans-and cis-isomers, respectively, demonstrating that the method is accurate within the desired range. Also, the results obtained from the analysis of pre-assayed creams spiked with different amounts of permethrin stock solution revealed acceptable recoveries with the mean values of 99.79% and 100.04% for trans- and cis-isomers, respectively, and . These values document a high recovery in this method.

3.3.4. Robustness

Chromatographic parameters including % of assay and permethrin isomers resolution were not significantly affected with the slight changes in the chromatographic conditions like alteration in flow rates, pH of the aqueous solution of mobile phase, and column temperature. Analysis was carried out in triplicate and only one parameter was changed in the experiments at a time. The results of the experimental variables evaluated were within the acceptable deviation , and the resolution of permethrin isomers were more than 2, indicating that the proposed method is robust under the conditions tested.

3.3.5. The Uncertainty of the Method

The expanded uncertainty of the method for quantification of permethrin isomers in pharmaceutical preparations was calculated. The results are shown in Table 3.

3.3.6. Application of the Method

The optimized and validated method was applied to the determination of permethrin isomers in marketed creams. The amounts of permethrin isomers in creams were calculated using calibration curve method. (Figure 4(a)) shows typical chromatogram obtained following the assay of Gilaranco creams. The results of the assay undertaken and the calculated uncertainties are shown in Table 3. The value of 99.77% of label claim indicates that the method is selective for the analysis of permethrin without interference from the excipients used to formulate and produce these creams.

4. Conclusion

The aim of this study was to develop a green and specific method for determination of permethrin isomers in pharmaceutical preparations. The method was designed to be specific, selective, sensitive, robust, reproducible, accurate, inexpensive, and easy to perform. The principal advantage of the method is the use of available environmentally friendly solvents and reagents for LC-analyzing and extractions to follow the first principle of green chemistry which emphasizes on waste prevention instead of remediation [18]. To the best of our knowledge, this is the first method which is thoroughly green and reports the metrological parameters in quantification of permethrin in topical preparations. In addition, recycling significantly reduced the mobile phase consumption and made the method economic. Moreover, the method is much more sensitive than the previously reported procedures [1216].

In conclusion, the newly developed method was successfully performed to simultaneous analysis of permethrin isomers in pharmaceutical preparations, and it can thus be used for routine analysis, quality control, and for studies of the stability of pharmaceutical formulations containing permethrin.

Conflict of Interests

The authors declared no conflict of interests.

Acknowledgments

The Authors extend their appreciation to Gilaranco Pharmaceutical Co. for kind donation of permethrin standard and raw material. This work was financially supported by Pharmaceutical Sciences Branch, Islamic Azad University (IAUPS).