Application and Validation of Simple Isocratic HPLC-UV-DAD Method with Dual Wavelength Detection for Ivabradine Determination and Its Application in the Study of Stress Degradation
Table 3
Validation parameters of proposed method.
Parameter
207 nm
286 nm
Linear work range [µg/ml]
5–100
5–100
Slope ± SD
1.3286 ± 0.110
0.2544 ± 0.001
Intercept ± SD
−0.3355 ± 0.277
0.0659 ± 0.012
Coefficient of correlation
0.9990
0.9993
Limit of detection (LOD) [µg/ml]
0.33
1.19
Limit of quantification (LOQ) [µg/ml]
1.09
3.97
Intra-assay precision 1 (% RSD, = 6; 10 µg/ml)
5.97
8.23
Intra-assay precision 2 (% RSD, = 6; 50 µg/ml)
5.81
1.60
Intra-assay precision 3 (% RSD, = 6; 100 µg/ml)
6.80
1.01
Interassay precision (% RSD, = 3; 10 µg/ml)
3.44
5.66
Recovery (%, )
92.1–102.2
95.6–109.5
Intra-assay precision concentrations were 10, 50, and 100 µg/ml, respectively.