Research Article

Rapid Separation of All Four Tocopherol Homologues in Selected Fruit Seeds via Supercritical Fluid Chromatography Using a Solid-Core C18 Column

Table 3

Intra- (n = 5 × 5) and interday (n = 25 × 5) recoveries, precisions, and accuracies of tocopherols extracted from apple seeds spiked at low, medium, and high levels and detected by the SFC-DAD system.

Recovery, precision, and accuracyLevels of tocopherol standardsTocopherols
α-Tβ-Tγ-Tδ-T

Intraday recovery (%)Low95948988
Medium98959794
High99999899

Interday recovery (%)Low96949191
Medium98979896
High991009998

Intraday precision (CV%)Low5.06.06.65.7
Medium2.62.42.82.9
High2.32.92.33.5

Interday precision (CV%)Low5.76.77.08.0
Medium2.73.23.34.0
High2.03.23.03.4

Intraday accuracy (bias%)Low4.66.510.911.8
Medium2.44.92.75.9
High1.20.81.70.6

Interday accuracy (bias%)Low4.46.09.39.4
Medium2.33.02.54.2