Review Article
Role of Electrophysiology in the Early Diagnosis and Follow-Up of Diabetic Retinopathy
Table 2
Standards for VEP assessment according to ISCEV.
| | Field size (deg) | Stimulus type | Stimulation | Background luminance (cdm−2) | Contrast (%) | Presentation rate |
| Pattern stimulation | >15 | Pattern reversal or onset/offset | Monocular | — | >75 | <1–3 reversals or ≤2 onsets per second | Flash stimulation | >20 | Standard luminance flash (2.7–3.4 cdsm−2) | Monocular (recommended) | 15–30 | — | <1.5 flashes per second |
|
|