Research Article
Cross-Camera External Validation for Artificial Intelligence Software in Diagnosis of Diabetic Retinopathy
Table 3
Sensitivity and specificity of VeriSee DR in diagnosing any DR in three different camera models.
| Fundus camera model | Sensitivity (95% CI) | Specificity (95% CI) |
| TRC-NW400 | 92.7% (87.5-97.9%) | 61.7% (56.8-66.6%) | TRC-NW8 series | 92.9% (88.4-97.4%) | 73.9% (69.2-78.6%) | Nomyd 8 series | 95.1% (92.0-98.2%) | 56.9% (51.2-62.7%) |
|
|