Research Article

Determination of Chromium in Natural Water by Adsorptive Stripping Voltammetry Using In Situ Bismuth Film Electrode

Table 5

Determination of the accuracy of the SqW-AdSV/in situ BiFE on the NASS 6a.

[NAAS 6]The content of chromium in the sample (ppb) ×1Chromium standard added (ppb)×oThe content of chromium in standard added samples (ppb) ×2Recovery (%)

2 ppb0.1161.8841.91596
2.043102
1.94197
Average ± S1.966 ± 0.054

6 ppb0.1165.8845.95899
6.258104
5,65594
Average ± S5.957 ± 0.213

10 ppb0.1169.88410.687107
10.125101
10.887109
Average ± S10.566 ± 0.279

aRecovery = (x2−x1).·100/x0; S is the standard deviation; experimental conditions are as shown in Table 3.