Two Different Maintenance Strategies in the Hospital Environment: Preventive Maintenance for Older Technology Devices and Predictive Maintenance for Newer High-Tech Devices

Table 5

Medical device failures.

Medical deviceTotal #NumberErrorsError code

Electrocardiogram502Not working ECG01
1Power circuit errorECG02
1Paper speed errorECG03
3Electrode errorECG04
1Sensitivity errorECG05

Pulse oximeter463Not workingSPO01
1Low BPM SPO02
1High BPMSPO03
2Low oxygen saturationSPO04
2High oxygen saturationSPO05
5Probe errorSPO06

Sphygmomanometer20012Not workingSPG01
15High pressure leakageSPG02
8Cuff errorSPG03
4Broken manometerSPG04
5Missing pieceSPG05

Infant incubator283Not workingINC01
2Over temperatureINC02
1Display errorINC03
1Baby probe errorINC04
2Broken coverINC05

Phototherapy unit303Not workingPHT01
2Low intensityPHT02

Defibrillator863Not workingDEF01
1Low/high energyDEF02
3Low batteryDEF03
2Electrode errorDEF04
1Paddle errorDEF05
2BPM errorDEF06
2Synchronization errorDEF07

Aspirator976Not workingASP01
4High vacuumASP02
3Low vacuumASP03
3Maximum vacuum ASP04
2Vacuum rise errorASP05

Electrosurgical unit521Not workingESU01
3Power circuit errorESU02
3High/low cut powerESU03
2High/low coag. powerESU04
1High/low bipolar powerESU05
2Foot switching errorESU06
2Patient electrode errorESU07