Research Article

Multichannel Surface EMG Decomposition Based on Measurement Correlation and LMMSE

Table 1

The number of reconstructed IPTs (Nr) (mean ± std. dev.), true positive rate (TPR) (mean ± std. dev.), and misplaced rate (MR) (mean ± std. dev.) for different decomposition methods.

MethodsParametersSNR (dB)
−10−50510

Measurement matrix autocorrelationNr8.2 ± 1.59.6 ± 0.69.6 ± 0.69.8 ± 0.59.8 ± 0.5
TPR (%)85.7 ± 1.997.6 ± 0.999.1 ± 0.399.4 ± 0.299.4 ± 0.5
MR(%)3.98 ± 1.062.16 ± 0.671.02 ± 0.230.98 ± 0.130.91 ± 0.06

GCKCNr5.0 ± 0.79.0 ± 1.210 ± 010 ± 010 ± 0
TPR (%)85.9 ± 4.099.5 ± 0.299.9 ± 0.199.6 ± 0.499.9 ± 0.0
MR(%)3.59 ± 1.121.05 ± 0.330.69 ± 0.260.66 ± 0.350.58 ± 0.17

MC-LMMSENr10 ± 010 ± 010 ± 010 ± 010 ± 0
TPR (%)92.8 ± 1.099.7 ± 0.0100 ± 0100 ± 0100 ± 0
MR(%)2.81 ± 0.851.02 ± 0.130 ± 00 ± 00 ± 0