A Novel Wire Is Effective for Echo-Guiding Percutaneous Atrial Septal Defect Closure: A Preclinical Study
Table 2
Procedural characteristics.
| Procedure characteristics | Study group () | Control group () |
| Total procedure time (min)# | 15.36 ± 4.86 | 25.82 ± 7.85 | Passing time (min)# | 2.69 ± 0.82 | 5.58 ± 3.34 | Frequency of the guide wire or delivery system dropping into the right atrium | 0/10 | 4/10 | Frequency of the guide wire traversing tricuspid valve | 1/10 | 2/10 | Frequency of atrial premature contraction# | 4.41 ± 2.61 | 9.60 ± 3.68 | Frequency of ventricular premature contraction# | 0.75 ± 0.36 | 1.34 ± 0.68 | Cardiac perforation | 0/10 | 0/10 | Pericardial effusion | 0/10 | 1/10 | Device implantation rate (%) | 100 | 100 |
|
|
Chi-squared test, ; #t-test, . |