Table 4

Maximum temperature in the thermal sensor next to the antenna slots for the computational model, phantom experimentation, and in vivo swine experimentation.

FEM modelPhantom experimentationIn vivo experimentationFEM model versus phantomFEM model versus in vivoPhantom versus in vivo

Power/time (W/s)Maximum temperature next to the slots (°C)ΔT (°C)
10/12092.1170.9189.3721.202.7418.46
20/40102.6069.1591.3133.4511.2922.16
30/30109.2076.7890.4232.4218.7813.64
50/15136.0075.22103.9660.7832.0428.74

Power/time (W/s)Maximum temperature at 1 cm (°C)ΔT (°C)
10/12069.3563.6072.885.753.539.28
20/4070.4059.8464.6110.565.794.77
30/3084.8567.1169.4517.7415.402.34
50/1580.6063.7185.2716.894.6721.56