VEMERS 2.0: Upgrading of an Emergency Use Ventilator from a Single Mandatory Volume Control Mode of Ventilation (VEMERS 1.0) to 8 Modes of Ventilation
Table 1
Working limits on VEMERS 2.0.
| Limit | Units | Value |
| Maximum cycle volume | ml | 800 | Minimum cycle volume | ml | 125 | Failsafe relief pressure | cm H20 | 60 | Maximum pressure in pressure control mode | cm H20 | 40 | Minimum pressure in pressure control mode | cm H20 10 | | Maximum peep | cm H20 | 25 | Highest trigger pressure in assisted mode | cm H20 | −3 | Lowest trigger pressure in assisted mode | cm H20 | −10 | Maximum respiratory frequency | bpm | 40 | Minimum respiratory frequency | bpm | 10 | Maximum IE ratio | | 3 | Minimum IE ratio | | 1 | Maximum flow | lpm | 60 | Maximum time apnea in pressure support mode | sec | 60 | Minimum time apnea in pressure support mode | sec | 5 | Maximum flow O2 therapy | lpm | 60 | Minimum flow O2 therapy | lpm | 10 |
|
|