VEMERS 2.0: Upgrading of an Emergency Use Ventilator from a Single Mandatory Volume Control Mode of Ventilation (VEMERS 1.0) to 8 Modes of Ventilation

Table 1

Working limits on VEMERS 2.0.

LimitUnitsValue

Maximum cycle volumeml800
Minimum cycle volumeml125
Failsafe relief pressurecm H2060
Maximum pressure in pressure control modecm H2040
Minimum pressure in pressure control modecm H20 10
Maximum peepcm H2025
Highest trigger pressure in assisted modecm H20−3
Lowest trigger pressure in assisted modecm H20−10
Maximum respiratory frequencybpm40
Minimum respiratory frequencybpm10
Maximum IE ratio3
Minimum IE ratio1
Maximum flowlpm60
Maximum time apnea in pressure support modesec60
Minimum time apnea in pressure support modesec5
Maximum flow O2 therapylpm60
Minimum flow O2 therapylpm10