Research Article

Highlights on Novel Technologies for the Detection of Antibodies to Ro60, Ro52, and SS-B

Table 2

Qualitative agreements between QUANTA Flash and BioPlex assays.
(a)

All patients ( )
BioPlex 2200 Ro52Percent agreement
(95% confidence interval)
PositiveNegativeTotal

QUANTA Flash Ro52
 Positive501161Pos agreement = 100.0% (92.9–100.0%)
 Negative0180180Neg agreement = 94.2% (89.9–97.1%)
 Total50191241Total agreement = 95.4% (92.0–97.7%)
kappa = 0.87 (95% CI 0.80–0.95)

(b)

All patients ( )BioPlex 2200 Ro60Percent agreement
(95% confidence interval)
PositiveNegativeTotal

QUANTA Flash Ro60
 Positive80080Pos agreement = 96.4% (89.8–99.2%)
 Negative3158161Neg agreement = 100.0% (97.7–100.0%)
 Total83158241Total agreement = 98.8% (96.4–99.7%)
kappa = 0.97 (95% CI 0.94–1.00)

(c)

All patients ( )BioPlex 2200 SS-BPercent agreement
(95% confidence interval)
PositiveNegativeTotal

QUANTA Flash SS-B
 Positive26329Pos agreement = 60.5% (44.4–75.0%)
 Negative17195212Neg agreement = 98.5% (95.6–99.7%)
 Total43198241Total agreement = 91.7% (87.5–94.9%)
kappa = 0.68 (95% CI 0.55–0.81)