Highlights on Novel Technologies for the Detection of Antibodies to Ro60, Ro52, and SS-B
Table 2
Qualitative agreements between QUANTA Flash and BioPlex assays.
(a) |
| All patients () | BioPlex 2200 Ro52 | Percent agreement (95% confidence interval) | Positive | Negative | Total |
| QUANTA Flash Ro52 | | | | | Positive | 50 | 11 | 61 | Pos agreement = 100.0% (92.9–100.0%) | Negative | 0 | 180 | 180 | Neg agreement = 94.2% (89.9–97.1%) | Total | 50 | 191 | 241 | Total agreement = 95.4% (92.0–97.7%) | | | | | kappa = 0.87 (95% CI 0.80–0.95) |
|
|
(b) |
| All patients () | BioPlex 2200 Ro60 | Percent agreement (95% confidence interval) | Positive | Negative | Total |
| QUANTA Flash Ro60 | | | | | Positive | 80 | 0 | 80 | Pos agreement = 96.4% (89.8–99.2%) | Negative | 3 | 158 | 161 | Neg agreement = 100.0% (97.7–100.0%) | Total | 83 | 158 | 241 | Total agreement = 98.8% (96.4–99.7%) | | | | | kappa = 0.97 (95% CI 0.94–1.00) |
|
|
(c) |
| All patients () | BioPlex 2200 SS-B | Percent agreement (95% confidence interval) | Positive | Negative | Total |
| QUANTA Flash SS-B | | | | | Positive | 26 | 3 | 29 | Pos agreement = 60.5% (44.4–75.0%) | Negative | 17 | 195 | 212 | Neg agreement = 98.5% (95.6–99.7%) | Total | 43 | 198 | 241 | Total agreement = 91.7% (87.5–94.9%) | | | | | kappa = 0.68 (95% CI 0.55–0.81) |
|
|