Effect of Using a Cardiac Catheterization Table-Stabilizing Stick on the Quality of Cardiopulmonary Resuscitation in the Cardiac Catheterization Laboratory: A Simulation-Based Study
Table 1
Effect of the catheterization table-stabilizing stick on the quality of cardiopulmonary resuscitation (CPR).
Table-stabilizing stick
Comparison
Use
Nonuse
Compression score (Q-CPR) (%)
79.6 ± 11.4
47.7 ± 30.3
31.9 ± 32.4
Depth (mm)
47.3 ± 2.9
40.8 ± 7.2
6.5 ± 7.7
Fully released chest compressions (%)
92.1 ± 8.3
96.3 ± 3.4
-4.1 ± 9.0
Sufficiently deep chest compressions (%)
36.4 ± 21.6
19.9 ± 17.3
16.6 ± 27.7
Table travel distance (mm)
0
6.6 ± 1.9
-6.6 ± 1.9
P < 0.01 for use and nonuse of table-stabilizing stick.